Kalibracyjne komory radonowe w Instytucie Fizyki Jądrowej im. Henryka Niewodniczańskiego
dc.contributor.author | Kozak, Krzysztof | |
dc.contributor.author | Janik, Mirosław | |
dc.contributor.author | Mazur, Jadwiga | |
dc.contributor.author | Kochowska, Elżbieta | |
dc.contributor.author | Bogacz, Joanna | |
dc.contributor.author | Łoskiewicz, Jerzy | |
dc.contributor.author | Swakoń, Jan | |
dc.contributor.author | Zdziarski, Tadeusz | |
dc.contributor.author | Haber, Ryszard | |
dc.date.accessioned | 2018-01-12T13:44:24Z | |
dc.date.available | 2018-01-12T13:44:24Z | |
dc.date.issued | 2003 | |
dc.description.abstract | The laboratory for calibration of radon detectors is presented in this report. The laboratory is equipped with two calibration chambers (volume: 200 and 600 liters), two radon sources (Ra-226 activity: 21,6 kBq and 52,3 kBq ), a pump, a flow meter, a manometer, etc. Range of obtained radon concentrations is from 35 kBq/m3 to 244 kBq/m3. The computer software has been elaborated for planning and controlling the calibration procedure. | pl_PL.UTF-8 |
dc.identifier.uri | http://rifj.ifj.edu.pl/handle/item/213 | |
dc.language.iso | pol | pl_PL.UTF-8 |
dc.publisher | Institute of Nuclear Physics Polish Academy of Sciences | pl_PL.UTF-8 |
dc.relation.ispartofseries | Raport IFJ PAN;1920/AP | |
dc.title | Kalibracyjne komory radonowe w Instytucie Fizyki Jądrowej im. Henryka Niewodniczańskiego | pl_PL.UTF-8 |