Repozytorium IFJ PAN

Dosimetric properties of the TLD European Cards with high sensitive MCP-7 (7LiF:Mg,Cu,P) thermoluminescence detectors for personal dosimetry

Pokaż uproszczony rekord

dc.contributor.author Bilski, Paweł
dc.contributor.author Budzanowski, Maciej
dc.date.accessioned 2018-01-12T11:18:28Z
dc.date.available 2018-01-12T11:18:28Z
dc.date.issued 2005
dc.identifier.uri http://rifj.ifj.edu.pl/handle/item/184
dc.description.abstract Dosimetric properties of Harshaw compatible cards equipped with two thermoluminescence MCP-7 (LiF:Mg,Cu,P) detectors were investigated at the accredited Laboratory of Individual and Environmental Dosimetry (Polish acronym LADIS) at the Institute of Nuclear Physics, Krakow. Dose detection threshold, dose sensitivity, dose linearity, residual signal, fading, sensitivity to thermal neutrons, sensitivity to ambient light, and the influence of environmental conditions were evaluated. All investigated dosimetric properties were found to be within the limits defined by IEC document 1066 “Thermoluminescence Dosimetry Systems for Personal and Environmental Standard”. pl_PL.UTF-8
dc.language.iso eng pl_PL.UTF-8
dc.publisher Institute of Nuclear Physics Polish Academy of Sciences pl_PL.UTF-8
dc.relation.ispartofseries Raport IFJ PAN;1964/D
dc.title Dosimetric properties of the TLD European Cards with high sensitive MCP-7 (7LiF:Mg,Cu,P) thermoluminescence detectors for personal dosimetry pl_PL.UTF-8
dc.type Report pl_PL.UTF-8


Pliki tej pozycji

Pozycja umieszczona jest w następujących kolekcjach