Repozytorium IFJ PAN

Optymalizacja elementów układu formowania wiązki dla stanowiska radioterapii protonowej w IFJ PAN metodami symulacji Monte Carlo

Pokaż uproszczony rekord

dc.contributor.author Dąbrowska, Joanna
dc.contributor.author Olko, Paweł
dc.contributor.author Sowa, Urszula
dc.contributor.author Swakoń, Jan
dc.date.accessioned 2018-01-12T09:08:27Z
dc.date.available 2018-01-12T09:08:27Z
dc.date.issued 2007
dc.identifier.uri http://rifj.ifj.edu.pl/handle/item/153
dc.description.abstract The work is aimed at optimizing the proton beam spreading system at the IFJ PAN radiotherapy of ocular melanoma facility. Single scattering foil and double scattering foil systems have been investigated by Monte Carlo simulations. In the case of single scattering foil system dose distributions were uniform to within ± 1 % over a 4-cm-diameter region, whereas in the case of double scattering foil system the same uniformity was achieved in a region up to 3.6 cm in diameter, but its efficiency of dose rates was 5.7 times better. Since the total thickness of foils in the single and the double foil systems was equivalent, in both cases the energy loss was the same. pl_PL.UTF-8
dc.language.iso pol pl_PL.UTF-8
dc.publisher Institute of Nuclear Physics Polish Academy of Sciences pl_PL.UTF-8
dc.relation.ispartofseries Raport IFJ PAN;1998/AP
dc.title Optymalizacja elementów układu formowania wiązki dla stanowiska radioterapii protonowej w IFJ PAN metodami symulacji Monte Carlo pl_PL.UTF-8
dc.type Report pl_PL.UTF-8


Pliki tej pozycji

Pozycja umieszczona jest w następujących kolekcjach

Pokaż uproszczony rekord