dc.description.abstract The time decay of the pulsed thermal neutron flux has been measured with quasi-point SOF detectors (scintillator with the optical fiber). The detectors have been placed at three positions in a 12 cm cube of polyethylene. The time decay constants of the fundamental exponential mode of the thermal neutron flux have been determined and compared with results of a reference measurement using 3He neutron detectors at the cube surface and with a theoretical result from the thermal neutron diffusion theory. pl_PL.UTF-8
dc.description.sponsorship The work has been partly sponsored by the Polish Ministry of Science and Higher Education under research within the Scientific Network “Neutrons–Emission–Detection” [Decision Nos. 620/E-77/SN-0055/07 and 620/E-77/BWSN-0055/2008] pl_PL.UTF-8